KingPo Technology Development Limited

IEC Testing Equipment

Manufacturer from China
Verified Supplier
9 Years
Home / Products / Test Finger Probe /

IEC 60601-1 Figure 7 Test Hook with 8mm Width and 1mm Thickness

Contact Now
KingPo Technology Development Limited
Visit Website
City:hong kong
Province/State:hong kong
Country/Region:china
Contact Person:MsLynette Wong
Contact Now

IEC 60601-1 Figure 7 Test Hook with 8mm Width and 1mm Thickness

Ask Latest Price
Video Channel
Brand Name :KingPo
Model Number :KP-TP68A
Certification :Calibration Certificate
Place of Origin :China
MOQ :1
Price :To be quoted
Payment Terms :T/T
Supply Ability :50 pcs per month
Delivery Time :7 working days
Packaging Details :safety carton pack or plywood box
Length :180mm
width :8mm
thickness :1mm
Angle :90°
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

IEC 60601-1-figure 7 Test Hook

IEC 60601-1 Figure 7 Test Hook with 8mm Width and 1mm Thickness

IEC 60601-1 Figure 7 Test Hook with 8mm Width and 1mm Thickness


TESTING AND MEASURING EQUIPMENT/ALLOWED SUBCONTRACTING

IEC 60601-1:2005 + Am.1:2012

Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

Clause Measurement/testing Testing / measuring equipment / material needed Subcontracting
4.11 Power input

Suitable devices for the voltage, current/power and frequency

Supply: 1 phase and 3 phase variacs

R
5.3 Ambient temperature, humidity, atmospheric pressure

Suitable devices for recording ambient temperature, humidity, atmospheric pressure

R
5.7 Humidity preconditioning treatment

Environmental conditions: Climate chamber controlling temperature and humidity

R
5.9.2 Accessible parts

Force gauge (30 N), standard test finger (figure 6), straight unjointed test finger, test hook (figure 7)

R
7.1.2 Legibility of markings

Illuminance meter

R
7.1.3 Durability of markings

Distilled water, ethanol (96% pure), isopropyl alcohol , timer / stop watch

R
8.4.2 Accessible parts including applied parts

Oscilloscope, oscilloscope leads, suitable instruments for measuring voltage, current, capacitance, test pin (Figure 8), metal test rod (D = 4 mm, L = 100 mm), force gauge (10 N)

R

8.4.3,

8.4.4

Limitations voltage and energy Suitable oscilloscope recorder /set-up & RCL meter R
8.5.5.1 Defibrillation protection

5 kV test circuit & oscilloscope interface circuit according to Figures 9 & 10, oscilloscope

S
8.5.5.2 Energy reduction test

The test circuit according to Figure 11, oscilloscope, oscilloscope leads

S
8.6.4 Impedance and current-carrying capability

Current source (25 A minimum, 50 or 60 Hz, 6 V maximum)

R
8.7 Leakage currents and patient auxiliary currents

Measuring device according to Figure 12, mains isolation transformers, variacs, voltmeter, millivoltmeter, aluminium foil, diverse circuits (fig. 13-20)

R
8.8.3 Dielectric strength

High voltage tester, isolating transformer for HV-tests (fig. 28), stop watch / timer

R
8.8.4.1 a) Ball pressure test

Test equipment according to IEC 60695-10-2

R
8.8.4.2 Resistance to environmental stress

Apparatus for ageing rubber in oxygen

S
8.9 Creepage distances and air clearances

Oscilloscope, oscilloscope leads, callipers, micrometer, spacing gauges, force gauge (2 N & 30 N), standard test finger (figure 6)

R
8.9.1.7 Material groups classification

Test equipment according to IEC 60112

S
8.9.3.4 Thermal cycling

Heating cabinet

R
8.11.3.5 Cord anchorage

Force gauge (at least 100 N), torque gauge (at least 0.35 Nm)

R
8.11.3.6 Cord guards

Weights, angle gauge, radius gauge

R
9.4 Instability hazards

5° & 10° inclined planes or inclinometer or trigonometric calculation, force gauge (at least 220 N), 20 cm by 20 cm test surface, weights, test threshold (10 mm high and 80 mm wide), 7 cm strap, stop watch / timer

R
9.5.2 Cathode ray tubes

The relevant tests of IEC 60065, Clause 18.

S
9.6.2.1 Audible acoustic energy

A-weighted sound pressure level according to ISO 3746, ISO 9614-1 or IEC 61672-1

S
9.6.3 Hand-transmitted vibration

Measurements are made in accordance with ISO 5349-1.

S
9.7.5 Pressure vessels

Hydraulic pressure test apparatus

S
9.8 Hazards associated with support systems

Weights or load cell, 0.1 m2 test surface, stop watch / timer, human body test mass (Figure 33)

R
10.1 X-radiation

Radiation meter

S
10.3 Microwave radiation

Radiation meter

S
10.4 Lasers Test equipment according to IEC 60825-1 S
11.1 Excessive temperatures

Temperature indicator/recorder suitable for this function and

thermocouples, 4 wire resistance unit, test corner, variac

R
11.2 Fire prevention

Spark ignition test apparatus (Figure 34), oxygen gas analyzer

S
11.3 Constructional requirements for fire enclosures FV tests specified in IEC 60695-11-10 S
11.6.2 Overflow

15° inclined plane or inclinometer or trigonometric calculation, stop watch / timer, high voltage tester

R
11.6.3 Spillage

Flask or graduated cylinder, stop watch / timer

R
11.6.5 Ingress of water or particulate matter

Classification tests of IEC 60529

W
11.6.6 Cleaning and disinfection

Dielectric strength and leakage current tests as appropriate

R
11.6.7 Sterilization

Sterilisation to client specification

S
13 Hazardous situations and fault conditions

stop watch / timer, voltmeter, ammeter, temperature indicator / recorder suitable for this function and thermocouples, 4 wire resistance unit, cheesecloth

R
15.3 Mechanical strength

Force gauge (250 N minimum), circular plane surface 30 mm in diameter, 500 g steel ball, 50 mm thick hardwood board (hardwood > 600 kg/m3), 40 mm step, hardwood doorframe (40 mm2), circulating air oven

R
15.4.2 Temperature and overload control devices

Positive temperature coefficient devices (PTC’s) with IEC 60730-1: 1999, clauses 15, 17, J.15 and J.17

S
15.4.3.4 Primary Lithium batteries

Performance of the tests identified in IEC 60086-4

S
15.4.3.4 Secondary Lithium batteries Performance of the tests identified in IEC 62133 S
15.4.6 Actuating parts of controls

Force gauge (at least 100 N), torque gauge (at least 6 Nm), stop watch / timer

R
15.4.7 Cord-connected hand-held and foot-operated control devices

Force gauge (minimum1 350 N), 30 mm diameter test tool, stop watch / timer

R
15.4.7.3 Entry of liquids

Classification tests of IEC 60529

W
15.5.1.1 Transformers

Winding tester for transformers, temperature indicator/recorder suitable for this function and thermocouples variac, loads

R
15.5.2 Dielectric strength

5x voltage / 5x frequency dielectric strength test fixture, stop watch / timer

R
16.6 Leakage currents

As in sub-clause 8.7

R
A 10.4 Light emitting diodes (LEDs)

As specified in IEC 62471

S
G Protection against hazards of ignition of flammable anaesthetic mixtures

4 mm & 12 mm diameter test rods, timer / stop watch, gasket test B-b of IEC 60068-2-2, force gauge (minimum 100 N), gas pressure meter (0 to 400 Pa range), voltmeter, ammeter, ohmmeter, meter for capacitance and inductance, oxygen mixture ( volume percentage 12,2 % ± 0,4 %),test apparatus for flammable mixtures (Figure G.7),

S
G.4.3 Prevention of electrostatic charges

antistatic material testing according to ISO 2882

S
L Insulated winding wires for use without interleaved insulation Sample is preparation according to IEC 60851-5:1996 & dielectric according to 60601-1; flexibility and adherence test 8 of IEC 60851-3:1996; mandrels of diameters according to Table L.1; heat shock test 9 of IEC 60851-6:1996; 2 mm diameter shot of stainless steel, nickel or nickel plated iron. S

Inquiry Cart 0